The eChamber is a small, battery-powered nebuliser you can use almost anywhere. It uses a vibrating mesh — a tiny plate covered in microscopic holes that shakes the liquid into a fine mist. It is quiet, quick and portable, but the mesh and the electrical contacts need looking after: keep the device dry, rinse the medication cup after every use, and give it a weekly vinegar-and-water clean. This page walks through using it and caring for it. It is for liquid medication nebulisation only.
What is in the box
- Lid × 1
- Mask × 2
- Device × 1
- Travel case × 1
- Medication cup × 2
- USB charging cable × 1
- Mouthpiece × 1
Before first use: clean, disinfect and dry all of the parts, and charge the device fully.
How to use it — step by step
Finishing and the weekly rinse
Then detach the medication cup by pushing the "press" button behind the device and pulling the cup upwards. Dispose of any leftover medication, and clean any accessories you used.
Cleaning
- Remove the medication cup, lid, and mouthpiece or mask from the device.
- Submerge and hand-wash these parts in lukewarm water (no more than 50 °C) for about 5 minutes.
- Set them aside to dry in a clean place, then disinfect if required.2
To disinfect (if needed): place the cleaned medication cup, lid and mouthpiece in a disinfectant solution (such as Milton, following its instructions), or steam-sterilise in a standard baby-steam device. Air-dry completely afterwards.
Charging & battery
- Charge with the supplied USB cable. If you use a mains adapter, make sure it complies with IEC 60601-1, 5.0 V.
- You can use the device while it is charging. A full charge takes about 1 hour.
- On low battery the light flashes slowly.
- Charge it at least once a month, even when it is not being used.
Looking after the mesh (medication cup)
The medication cup is a semi-disposable consumable with a maximum warranty of three months. Because vibrating mesh pushes liquid through microscopic pores, a fine mineral film gradually builds up and narrows them over time — this happens with every vibrating-mesh nebuliser and is not a fault3. To get the longest life from each cup:
- Rinse the cup thoroughly with fresh water after every use.
- Where possible use distilled (demineralised) water for rinsing — it is largely mineral-free and prevents build-up.
- Clean and dry the cup after every use to minimise residue.
- Be aware that hypertonic saline (salt water) shortens cup life by speeding up mineral deposits; non-salt medicines are gentler on the mesh.
Keep the pins and plates clean
The device works by making contact between the connector pins on the medication cup and the metal plates on the body. If either becomes dirty or corroded, the nebuliser may light up but produce no mist, because the signal cannot pass. After each use, gently clean and dry both the pins and the plates, leave no moisture or medicine on them, and store the device somewhere clean and dry.
Troubleshooting
| Problem | Likely cause | What to do |
|---|---|---|
| Won't turn on / no light | Device isn't switched on | Turn it on and check the light comes on. |
| Button is jammed | Press the ON/OFF switch to dislodge it. | |
| Device isn't charging | Charge it (see Charging & battery). | |
| USB cable may be broken | Use a cable with the same voltage/amp rating, or contact warranty. A non-matching cable may void warranty. | |
| Mist is too weak | Not enough medicine in the cup | Refill — no higher than the 10 ml line. |
| Wrong medication in use | Check with your prescriber; see product warnings. | |
| The spray mesh is blocked | Put 3 drops of white vinegar and 3–6 ml water in the cup and run it until the cup is empty — do not inhale this — then clean the cup. | |
| Lights up but no mist at all | Connector pins or metal plates are dirty or corroded | Switch off, then gently clean and dry both the pins on the cup and the plates on the body (see Keep the pins and plates clean). |
| Some medicine left in the cup | This is normal | A small residual amount is expected. |
| Strange noise | Possible product fault | Contact warranty / Bird Healthcare. |
Storage
Store the device in its carry case, keeping the temperature between −20 °C and 55 °C, relative humidity under 80%, and atmospheric pressure between 70 and 106 kPa. Do not leave it unused in storage for more than 5 years.
Bird Healthcare will voluntarily replace any e-chamber nebuliser made in 2021 or earlier (the manufacture date is printed on the base) with the current model at no cost — the newer connector alloy resists corrosion and conducts better. To arrange a replacement, email service@birdhealthcare.com.
References & evidence base
- Bird Healthcare. e-chamber nebuliser (BHC-NEBPRO-1) — instructions for use, cleaning and disinfection. Melbourne: Bird Healthcare; 2026.
- Boe J, Dennis JH, O’Driscoll BR, et al. European Respiratory Society guidelines on the use of nebulizers. Eur Respir J 2001;18(1):228–242.
- Ari A, Restrepo RD; American Association for Respiratory Care. Aerosol delivery device selection for spontaneously breathing patients. Respir Care 2012;57(4):613–626.
References are numbered in citation order (Vancouver/BMJ style) and were current at the time of writing. Cleaning and maintenance steps follow the manufacturer’s instructions for this device — always follow the leaflet supplied with your own nebuliser and your prescriber’s advice.
More than one of our services applies here, and which combination suits you depends on what your assessment shows.
Corrections: If something on this page is wrong, out of date or unclear, we want to know. Email reception@inspireclinic.au with the page name and what you believe is incorrect. Substantive corrections are made promptly, and the guide’s version and last-updated date are changed to reflect it.